You are actually in BAMO EN » Our Products » Old Docs » pH - O.R.P. & Chlore » CP2.1
General Catalog 193-05
CP2.1
CP2.1
Total chlorine sensor
CP2.1
Warning! You are consulting an archived documentation:

This product is not anymore manufactured or marketed by BAMO
Do not hesitate to contact us if you are looking for a substitution product certified or if you have a question


Complete system (Assembly sold separately)
- Water treatment, drinkable quality and sea water
- Low incidence of pH changes
- Surfactants: partial tolerance
- Operating temperature 45 °C max.
- Operating pressure 0.5 bar max.
TECHNICAL FEATURES
Measuring parameter
:
Total chlorine (Free + combined chlorine)
Applications
:
Water treatment (drinkable quality water and sea water)
Surfactants: partial tolerance
Chlorination additives
:
Inorganic compounds, such as NaOCl - Ca(OCL)₂ -
gaseous chlorine - chlorine by electrolysis
Measuring system
:
Closed cell with 3 electrodes and electrolyte
Output signal
:
4 ... 20mA, screw connectors (2 x 1 mm²)
No galvanic insulation
Operating temperature
:
1 to 45 °C
Automatic temperature compensation
Operating pressure
:
0.5 bar max. (No vibrations, no pulsating flow)
Flow rate
:
About 30 l/h
Supply voltage
:
12 ... 30 V DC, Load = 500 to 900Ω
pH operating range
:
pH 4 to pH 12 (low incidence of pH)
Interferences
:
Incidence of ClO₂ and O₃ on measuring signal
Zero adjustment
:
Not necessary (from factory)
Slope calibration
:
Only 1 point with BAMOPHAR 194 according to directive DPD-4
Materials
:
PVC-U, PEEK, AISI 316 Ti
Dimensions
:
O.D. 25mm, length 220mm (4-20mA)

 

CE Conformity: The instrument meets the legal requirements of the current
European Directives
CODE NUMBERS AND REFERENCES
Code
Reference
Measuring range
Resolution
193 043
CP2.1.MA2
0.01 ... 2 mg/l
0.1 mg/l
193 044
CP2.1.MA5
0.01 ... 5 mg/l
193 045
CP2.1.MA10
0.01 ... 10 mg/l
Spare parts
193 903
M48.2
Sensor end with diaphragm for CP2.1MA sensor
193 XXX
ECP1.4/GEL
Electrolyte for CP2.1MA
 
Other versions on request (measuring range, power supply, output signal, connector, etc.)
 
Precautions
Flow rate must be constant and a measuring cell with flow controller is necessary (see data-sheet 193-95).
In order to install easily a complete system, we propose assemblies designed for specific applications (on request).
General Catalog 193-05
21/09/2022 11:42
Total chlorine sensor
CP2.1
Technical & Commercial support's main contacts
Ask a QUOTATION online
BAMO Mesures SAS
22, Rue de la Voie des Bans
Parc d'activités de la Gare
95100 ARGENTEUIL France
Tel: +33 1 30 25 83 20
Fax: +33 1 34 10 16 05
E-mail: export@bamo.fr
URL: http://www.bamo.fr
Siret: 314 055 864 000 65
TVA Intra: FR 08 314 055 864
APE: 2651 B
Capital: 400 000 Euros
Built by CARIMEDIA since 1998
© 1998-2024 All rights reserved - Disclaimer