For swimming pool water, drinkable water, cooling water, process water
Range: from 0.001 up to 10.0 g/L
Output signal: 4-20 mA
Sensor affords surfactants
Pressure limit: 1 bar as a maximum
Unnecessary zero adjustment
DESCRIPTION
Principle Amperometric method with a closed cell of 2 electrodes; wasted sample. The probe includes a CTN sensor for the temperature compensation.
Mounting / Recommendations The measuring at a constant flow rate requires the use of a specific cell holder (data sheet 193-95). The complete assembly optimizes the operations.
TECHNICAL FEATURES
Range: 0.001 to 10.0 g/L pH range: From 2 to 11 Operating pressure: 1 bar as a maximum Operating temperature: 1 to 45°C Flow rate limits: 30 to 40 L/h (see doc. 193-95) Power supply: 12...30 V DC, [Rmax = (U-7,5) / 20 kΩ]
Materials: PVC-U, electro-polished AISI 316L DimensionsØ 25 mm, length 225 mm
General Catalog 193-25 17/04/2018 10:36
Hydrogen peroxide sensor WP7
BAMO Mesures SAS
22, Rue de la Voie des Bans Parc d'activités de la Gare
95100 ARGENTEUIL France