The TRUBOMAT measures the turbidity of liquids using the light absorption and scattered light measuring method. The inline measuring cell includes two IR emitters and one IR receiver in pulsating light mode (IR measuring method according ISO 7027).When interconnected, the BAMOPHOX TUR, measuring amplifier, analyses the 4 to 20 mA signal generated by the TRUBOMAT.
DESCRIPTION
The flow-through cell allows the turbidity measurement directly on the line, without sampling system. Its built-in transmitter is convenient for transmitting the signal to installed DCS or PLC or BAMOPHOX TUR. The 4-20 mA corresponds to the DIP Switch selectable range:
TRUBOMAT 20: from 0.01 to 20 FNU in 5 selectable ranges
TRUBOMAT 1000: from 0.1 to 1000 FNU / FAU in 5 selectable ranges
APPLICATIONS
• Drinkable and waste water plants, well water monitoring • Filtration control, flocculants dosage, phase separation • Pollution survey
TECHNICAL FEATURES
Power supply: 24 V DC rated (10...30 V DC)
Power consumption: 0,5 W
Output signal: 4-20 mA, for connection to the BAMOPHOX TUR 436
Ambient temperature: +5...+45 °C
Process temperature: +5...+100 °C
Material: 316L Stainless steel body / Terminal housing plastic PBT, IP 65
Accuracy: ±5% reading, ±1% full scale in use
Resolution: 0,001...0,2 FNU according to measuring range
Status indication: Green LED = correct Red LED = failure
Selectable ranges TRUBOMAT 1000: 0.1 to 50 FNU 0.1 to 100 0.1 to 200 0.1 to 500 (or 0.1 to 300*) FNU 100 to 1000 FNU * fittings DN80 & DN100
Selectable ranges TRUBOMAT 20: 0.01 to 1 FNU 0.01 to 2 FNU 0.01 to 5 FNU 0.01 to 10 FNU 0.01 to 20 FNU * fittings DN80 & DN100
Material: Stainless steel body
Process connections: PN10 flanges (DIN 2633), Triclamp (DIN 11850)
CE Mark: In accordance with low-voltage directive (2006/95/EG), EMC directives (2004/108/CEE)
General Catalog 446-01 17/04/2018 10:58
In line turbidity transmitter TRUBOMAT® GAB
BAMO Mesures SAS
22, Rue de la Voie des Bans Parc d'activités de la Gare
95100 ARGENTEUIL France