Principle Free chlorine measurement by amperometric method with a diaphragm cell of 2 electrodes; wasted sample. The probe includes a CTN sensor for the temperature compensation.
Mounting / Recommendations The measuring at a constant flow rate requires the use of a specific cell (see data sheet 193-95). The complete assembly optimizes the operations
CAUTION
The sample may not content surfactants and the pH must be lower than 8
TECHNICAL FEATURES
Range: 0.5 to 200 ppm (free chlorine) Measuring signal depends of pH; regulation of pH is necessary. Operating pH range: From 6 to 8 and regulated Pressure limit: 1 bar as a maximum Temperature limits: From 1 to 45°C Flow rate limits: From 30 to 40 L/h (see the data sheet 193-95) Power supply: 12 ... 30 V DC, [Rmax = (U-7,5) / 20 kOhm] Materials: PVC-U Dimensions: Ø 25 mm, length 225 mm
General Catalog 193-02 17/04/2018 10:32
Free chlorine sensor, at constant pH CL4.1
BAMO Mesures SAS
22, Rue de la Voie des Bans Parc d'activités de la Gare
95100 ARGENTEUIL France