You are actually in BAMO EN » Our Products » Old Docs » Oxygen » BAMOX 15
General Catalog 460-01
BAMOX 15
BAMOX 15
O2 Measure electrode
BAMOX 15
Warning! You are consulting an archived documentation:

This product is not anymore manufactured or marketed by BAMO
Do not hesitate to contact us if you are looking for a substitution product certified or if you have a question



- Robust construction.
- Large surface membran.
- Easy and quick replacement interchangeable and regeneratable cartridge.
- Galvanic measuring system with membran.
- Silver or Gold cathode / Lead anode.
- Continuous filling up by KHCO3 gel.
- Very shirt time reply.
- Self cleaning system for stagnant water.
- Maximum depth using 100 m (Water column)
DESCRIPTION
Those electrodes are dedicated for wast water treatement process or aquaculture. Measure element (Lead anode and Silver or Gold cathode) is interchangeable. This one can be regenerated by electrolyte gel and Lead anode replacement, for low cost.
In general case, Silver cathode is commonly used, Gold cathode is mainly use for sulphurous water.
Fixing systems for circulation, immersion or floating immersion measure are available
TECHNICAL DATA
Cathode: Silver BAMOX 15.1
Gold BAMOX 15.2
Anode: Lead
Electrolyte: KHCO3 gel
Does not to be change in case of membran replacement.
Anode and cathode are set on cartridge
- Electrolyte regeneration and Lead anode -.
Diaphragm: Polyethylene 0,025 mm
Reply time: 35 s for 0 - 90% value changement.
Flow speed: 0,5 m/s mini
Temperature: 5 to 35°C
Max. pressure: 100 m (Water column)
Cable: Standard 5 m (others on request - 100 m maxi)
Installation: PVC circulation holder -ND25 M to be stick. Immersion probe 2 metre length (others on request). Probe holder page 143
OPTION
Self cleaning system for to ensure correct measure in stagnant water.
Direct fixing on BAMOX 15.1 or BAMOX 15.2 electrodes.

Supply: Standard 24Vdc - 50 mA
12 Vdc - 100 mA or 6 Vdc - 200 mA on request.
Rotation speed : 10 tr/min by integrated reduicer.
Max. pressure: 35 m (water column)
Materials: PVC frame - Nylon brush and cover
General Catalog 460-01
28/10/2010 08:59
O2 Measure electrode
BAMOX 15
Technical & Commercial support's main contacts
Ask a QUOTATION online
BAMO Mesures SAS
22, Rue de la Voie des Bans
Parc d'activités de la Gare
95100 ARGENTEUIL France
Tel: +33 1 30 25 83 20
Fax: +33 1 34 10 16 05
E-mail: export@bamo.fr
URL: http://www.bamo.fr
Siret: 314 055 864 000 65
TVA Intra: FR 08 314 055 864
APE: 2651 B
Capital: 400 000 Euros
Built by CARIMEDIA since 1998
© 1998-2024 All rights reserved - Disclaimer