For drinkable water monitoring
Range: from 0.01 up to 10 ppm
Output signal: 4-20 mA
Pressure limit: 0.5 bar as a maximum
Unnecessary zero adjustment
DESCRIPTION
Principle Free chlorine measurement by amperometric method with a diaphragm cell of 3 electrodes; wasted sample. The probe includes a CTN sensor for the temperature compensation.
Mounting / Recommendations The measuring at a constant flow rate requires the use of a specific cell (see data sheet 193-95). The complete assembly optimizes the operations.
Note: The water sample may not content surface-active additives.
TECHNICAL FEATURES
Range: 0.01 to 10 ppm (free chlorine) pH changes do not affect the measure within pH values 4 to 11
Pressure limit: 0.5 bar as a maximum Temperature limits: From 1 to 45°C Flow rate limits: From 30 to 40 L/h (see the data sheet 193-95) Power supply: 12 ... 30 V DC, [Rmax = (U-7,5) / 20 kΩ] Materials: PVC-U, electro-polished AISI 316L Dimensions: Ø 25 mm, length 225 mm
General Catalog 193-04 20/11/2018 16:30
Free chlorine sensor, without effect from pH changes CS2.3
BAMO Mesures SAS
22, Rue de la Voie des Bans Parc d'activités de la Gare
95100 ARGENTEUIL France