Warning! You are consulting an archived documentation:
This product is not anymore manufactured or marketed by BAMO
Do not hesitate to contact us if you are looking for a substitution product certified or if you have a question
-2 Switchable scales: 0 - 2 M? (Centred 100 K?) 0 - 20 M? (Centred 1 M?) -1 Adjustable set point -Volt free change over relay -Delated analogic 4-20 mA output
TECHNICAL SPECIFICATIONS
Main power supply:230 V/ 50-60 Hz 10 VA single phase. Consumption: 10 VA Dilated scales:0 - 2 M? (Centred 100 K?) 0 - 20 M? (Centred 1 K?) Probe : Coefficient C=0,1 (C=1 on request) Accuracy:± 3% Set-points:adjustable with potentiometer and push-button Output relay:Volt free change over contact rated at 500VA / 220V / 3A Indication with green and red LED Analog output:4 to 20 mA (max. 300 ?) Housing:DIN 72 x 144 mm panel mounting IP 20. Connections with removable wiring connectors. Wall mounting IP 55. Connections with 9 mm cable glands
ASSOCIATED PROBES
IN LINE MEASURE 360 315BS 1285Design: 316 Stainless Steel / PTFE - Connection 1/2" M.G. connection aluminium head 360 500BC 1425Design: 316 Stainless Steel / PTFE - Connection 1/2" M.G. connection aluminium head Supplied with 3 M coaxial wire LINK WIRE 368 100 CCA wireSpecial wire for conductivity or resistivity measure.
OPERATING INSTRUCTIONS
The resistivity meter is factory calibrated. The Set - Point adjustment is made with the potentiometer located in the front face. The probe wires must be separated from any power wires as all measuring wires. Use only the wire type CCA.
SET POINT ADJUSTMENT
Press the membrane switch located on the front, the needle indicates the Set - Point, with the potentiometer adjust the required value. The relay will switch at this setting, the green LED is on when the measure is higher then the preset value, the LED turns red when the measure is lower then the preset value.
GENERAL CHECK
With a short circuit instead of the probe the meter indicates 0. When the circuit is open the measure is infinite. A x? resistance provide a 10 x? indication. Example: Resistance of 100 K? instead of the probe, the meter indicates 1 M?.
General Catalog 310-01 28/10/2010 08:57
Resistivity-meter 801 ED - 801 MD
BAMO Mesures SAS
22, Rue de la Voie des Bans Parc d'activités de la Gare
95100 ARGENTEUIL France