Search
MENU Catalog Global Sites
BAMO
FRANCE
French manufactured of sensors and control instruments for fluids
You are actually in BAMO EN » Our Products » Old Docs » Conductivity » SERIES LAS
General Catalog 315-01
SERIES LAS
SERIES LAS
Resistivity meter indicator + sound alarm
SERIES LAS
Warning! You are consulting an archived documentation:

This product is not anymore manufactured or marketed by BAMO
Do not hesitate to contact us if you are looking for a substitution product certified or if you have a question

-Measuring range 0 to 2 M?
-Indicator with mechanical meter
-Adjustable sound alarm
-Compact housing for wall mounting
-2 LED indicators Green / Red
-Associated probe: BC 1425 / BNC
FUNCTIONING
The meter indicates permanent the resistivity.
Alarm preset is done internal with a push-button and potentiometer.
In case the indication is below the preset alarm value a buzzer will sound and the red LED turns on until the resistivity meter reaches the correct value again.
The meter has a expanded 0-2 M? scale that insures a good reading from 20 to 500 K?. The centre indicates 100K?.
The LAS is perfectly suitable for demineralized applications where there is no need automatic from alarm.
ASSOCIATED PROBES : BC 1425 / BNC
All electrodes with coefficient 0.1 are adaptable to this type of meter (See note DOC 360).
However for economical reasons the particularly adapted probe is the probe BC 1425/ BNC-R 1/ 4 delivered with 5 meter coax cable and BNC connector to insure simple installation and service.
TECHNICAL SPECIFICATIONS
Measuring range:0 - 2 M? / centre 100 K?
Alarm:Sound buzzer and red LED indicator
Housing:Macralon 80 x 160 x 55 mm, fixation with 4 holes
Power supply:230 VAC / 50 Hz
Probe:BC 1425 BNC / - R 1/4" Mg - C=0.1
delivered with 5 meter cable and BNC connector
STARTING UP
1) - The probe must always be on the fluid
2) - Connect the LAS on the main supply and the probe on the plug provided for.
3) - To calibrate the set, take off the cover, push the press-button and screw the potentiometer. The arrow show you the preset value.
VERIFICATION
1) Testing of the LAS:
Short-circuit the BNC adapter on the meter, the meter has to indicate 0.
The red LED lights and the buzzer alarm sounds.
Without connection of the probe the meter reads more then 2 M?.
With a pure resistance of 10 K? connected instead of the probe the meter shows 100K?.
General Catalog 315-01
28/10/2010 08:57
Resistivity meter indicator + sound alarm
SERIES LAS
Technical & Commercial support's main contacts
Ask a QUOTATION online
BAMO Mesures SAS
22, Rue de la Voie des Bans
Parc d'activités de la Gare
95100 ARGENTEUIL France
Tel: +33 1 30 25 83 20
Fax: +33 1 34 10 16 05
E-mail: export@bamo.fr
URL: http://www.bamo.fr
Siret: 314 055 864 000 65
TVA Intra: FR 08 314 055 864
APE: 2651 B
Capital: 400 000 Euros
Built by CARIMEDIA since 1998
© 1998-2024 All rights reserved - Disclaimer