Search
MENU Catalog Global Sites
BAMO
FRANCE
French manufactured of sensors and control instruments for fluids
You are actually in BAMO EN » Our Products » Old Docs » Conductivity » SERIES 319 ML-DB
General Catalog 319-01
SERIES 319 ML-DB
SERIES 319 ML-DB
Resistivity meter
SERIES 319 ML-DB
Warning! You are consulting an archived documentation:

This product is not anymore manufactured or marketed by BAMO
Do not hesitate to contact us if you are looking for a substitution product certified or if you have a question


- 2 independent measuring inputs
- For coefficient probe C= 0.1
- Wall mounting enclosure IP 55
SYSTEM 1
- Resistivity meter blind
- Adjustable alarm with internal potentiometer
- Volt free change-over output relay
SYSTEM 2
- Resistivity meter with LCD display
- Analog 4 - 20 mA linear output
- Scale 0-20 M Ohm standard
DESCRIPTION
BAMO developed the resistivity meter 319 ML / DB to control and supervise two bottles of resins meadow to mixed bed climbings in series.
SYSTEM 1: Controls the first bottle in blind mode.
As soon as the resistivity is inferior to the preset value adjusted with the internal potentiometer than relay 1 switches (preset manufacture value = 500K?).
The change-over relay S1 changes state, meaning for the user that the first bottle is exhausted and that it would be desirable to change it before the second bottle (that functions again) is exhausted.
SYSTEM 2: Controls bottle two in the same manner but to an other meadow value-ruled by the user (Threshold and relay S2).
Furthermore, the resistivity at the output of the second bottle is permanent displayed.
The 4-20 mA output signal allows to record and report the second system from a distance and gives an impression of the water quality at the output of the two bottles.
TECHNICAL SPECIFICATIONS
Power:230 VAC/ 50 Hz
Power consumption:5 VA
Measuring scales:0-20 MOhm on system 2 (others on request)
2 MOhm on system 1
Precision:±1%
Preset system 1:With Internal potentiometer
Preset system 2:With external potentiometer and push-button,
value displayed on the LCD display
Relay:2 contacts Volt free change-over
PC 3 A/ 250 V/ 500 VA
Analog output:2-wire 4-20 mA (not galvanic isolated)
Housing:Wall mounting IP 55 with transparent lid
Connection:Screw terminals with separated power connections.
Outputs with PE cable gland
Probes:Coefficient C= 0.1.
CONNECTIONS
The green LED in the front indicates that the output relay is in rest position.
2 independent high or low alarm thresholds are presetable with the switches located at the front.
The analog output is proportional to the beach of measure of second bottle
(0 or 4-20 mA for 0 to 100% of the scale). The output can be programmed with dip-switches located behind the front as 0-20 or 4-20 mA.
CALIBRATION
The resistivity meters are deliverd factory calibrated according to the characteristics required with the order.
In case of scale modification on the analog output return the unit to the factory.
Our technical service is at your disposition for all complementary information.
FRONT PCB INFORMATION
The potentiometers and dip-switches are lokated behind the front. Remove the four screws in the front to acces the PCB.
General Catalog 319-01
28/10/2010 08:58
Resistivity meter
SERIES 319 ML-DB
Technical & Commercial support's main contacts
Ask a QUOTATION online
BAMO Mesures SAS
22, Rue de la Voie des Bans
Parc d'activités de la Gare
95100 ARGENTEUIL France
Tel: +33 1 30 25 83 20
Fax: +33 1 34 10 16 05
E-mail: export@bamo.fr
URL: http://www.bamo.fr
Siret: 314 055 864 000 65
TVA Intra: FR 08 314 055 864
APE: 2651 B
Capital: 400 000 Euros
Built by CARIMEDIA since 1998
© 1998-2024 All rights reserved - Disclaimer